Triggered Samples Count、Last Number of Samples

LASAIR III Operations Manual ​

https://manualzz.com/doc/9978640/lasair_iii_operations_manual#p80

Triggered Samples Count (n)
Allows entry of the number of samples (1 to 35) that should have exceeded the alarm limit value to turn on the alarm. The value of this field should be less or equal to the Last Number of Samples value (m).

Last Number of Samples (m)
Allows entry of the time frame (1 to 35) windows for the alarm. The value in the field shows how many last samples will be used for determining alarm condition. The value of this field should be grater or equal to the Triggered Samples Count value (n). If the Enable Flag is set and Last Number of Samples value (m) is less than Triggered Samples Count value (n), then (n) is set equal (m). The label for the column is (m) string. The default value is 1.

1/0
A toggle button to turn the Enable Flag On (a check mark displays) or Off (a blank button displays).The Enable Flag enables using the alarm for the channel.

Alarm Condition
The alarm goes on if there are at least n samples from last m samples where the alarm limit value exceeded. The following table shows an example of the state of alarm after different samples with an alarm limit value is 1000.

f:id:ti-nspire:20211119150853p:plain:w300

非公式訳:
Triggered Samples Count (n)
サンプルが何回アラームリミット値を超えたらアラームをオンにするのかが入力できます。1~35の範囲で入力できます。このフィールドの値はLast Number of Samples値(m)以下にしてください。

Last Number of Samples (m)
アラームの対象期間をサンプル数で入力できます。1~35の範囲で入力できます。アラームを発生させるかどうかを過去何回ぶんのサンプルで判断するのかが入力できます。このフィールドの値は、Triggered Samples Count値(n)以上にしてください。Enable Flagが立っていて、かつLast Number of Samples値(m)がTriggered Samples Count値(n)よりも小さい場合、(n)は(m)にセットされます。この列のラベルは(m)文字列です。デフォルト値は1です。

1/0
Enable Flagをオン(チェックマークが表示される)またはオフ(チェックマークが消える)にするトグルボタンです。Enable Flagが立っていれば、そのチャンネルのアラームが使えます。

アラーム条件
過去m回のサンプルのうち、アラームリミット値を超えたサンプルがn回以上あった場合にアラームがオンになります。以下の表は、複数のサンプルを取得したあとのアラームの状態を例示したものです。アラームリミット値は1000に設定してあります。